Most recently this same group published an abstract validating transmural myocardial perfusion ratios (at rest) using 256-slice MDCT in a group of patients as compared with SPECT perfusion imaging.
Methods for Myocardial Perfusion
In order to determine contrast timing in the systemic circulation using MDCT it is common to perform a ‘test bolus’ where the contrast clearance characteristics are determined, most commonly in the aorta. Essentially this is a ‘perfusion’ curve for the aorta. By also acquiring indicator-dilution curves over a region of interest in the myocardium and comparing the data (equation 1), an estimation of myocardial perfusion (flow per unit volume) can be defined.
Thus, in principle, myocardial perfusion imaging is now possible using the currently installed 64+-slice MDCT scanners throughout the world. However, this is a bit more difficult to put into standard clinical practice due to a variety of logistical issues as well as the potential for significant radiation exposure to the patient.
In principle the methods could be summarized as follows:
- Perform ‘whole heart’ CCTA every few seconds for 10+ exposures beginning prior to intravenous ‘bolus’ injection of iodinated contrast and past the pre-determined ‘circulation time’ – so as to provide information to develop indicator dilution curve characteristics in the aorta/left ventricular cavity and myocardium. Such studies would be best done at end-systole rather than end-diastole to maximize the thickness of the myocardium
- This procedure should be repeated during continuous infusion of adenosine or possibly following a bolus injection of an A2A-adenosine receptor agonist.
- Perform post-processing of the indicator dilution curves from the aorta and any operator defined myocardial region of interest and apply the subsequent data using Equation 1.
Technical Issues and Radiation
The major obstacle to performing myocardial perfusion imaging using MDCT relates to radiation exposure to the patient. Using retrospective gating methods to perform CCTA, the effective radiation to the patient ranges from 6-15 mSv. If one were to then repeat this ‘resting’ study with another similar study during adenosine infusion of bolus injection of an adenosine receptor agonist – the total radiation dose would at least be doubled.
However, the introduction of newer prospective gating protocols for CCTA reduce radiation dose considerably, generally to below 5 mSv. Performing two such studies would then result in generally no more radiation exposure than an original retrospective examination.
The introduction of 256-slice MDCT (Phillips) and 320-slice MDCT (Toshiba) allows imaging of the entire heart without the requisite spiral overlap required for similar imaging using 64-slice CT. This then can result in even further reduction of radiation exposure to perhaps one-quarter to one-fifth of that of any other 64-slice examination. ‘Whole Heart’ perfusion imaging could potentially be done with 256- 320-slice MDCT (perhaps with also providing information on regional epicardial coronary anatomy) at a single setting with rest and ‘stress’ images for radiation doses significantly below that possible with standard Technecium-99 SPECT imaging (currently at 10-15 mSv).
Conclusions
The promise of defining regional myocardial perfusion using MDCT could potentially complete the long term efforts to have a practical, clinical, non-invasive imaging method which provides details on cardiac size, cardiac function, coronary anatomy, and coronary blood flow; the recent development of methods to use MDCT effectively but reduce the overall radiation exposure to the patient and/or provide simultaneous whole heart imaging make this goal potentially achievable.
Of course, as with all such investigations, additional validation will be necessary and more practical methods of implementation and advanced image processing must be found prior to allow wide-spread application – but the future is indeed most promising. |